#1 in Distributing Specialty
Chemicals & Equipment

Email Us Today Free shipping on
all chemicals (48 states)

  Home > Test Kits, Reagents, Etc. > Taylor Test Kits >

Taylor K-1575
Taylor K-1575
Alkalinity Test Kit


 
Taylor K-1575
Our Price: $97.99
Sale Price: $84.99
Savings: $13.00

Quantity in Stock:100


Availability: Usually Ships in 3 to 5 Business Days
Product Code: K-1575
Qty:

Description Shipping & Transit Times
 

Taylor K-1575

ChemWorld is a distributor of Taylor Technologies Test Kits and Reagents. Order directly online and save today.

DROP TEST P/M & P/T ALKALINITY (1 drop = 10 or 50 ppm)

COMPONENTS: 1 x 5085 Instruction 1 x 9198G Sample Tube, Graduated, 25 mL, plastic w/cap and green dot 1 x R-0637-C Methyl Orange Indicator, 2 oz, DB 1 x R-0638G-C Phenolphthalein Indicator (green cap), 2 oz, DB 1 x R-0645-C Total Alkalinity Indicator, 2 oz, DB 1 x R-0724-C Hydrochloric Acid .12N, 2 oz, DB 1 x R-0735G-C Hydrochloric Acid .6N (green cap), 2 oz, DB.

P/M Alkalinity

NOTE:

For 1 drop = 10 ppm, use R-0724 Hydrochloric Acid .12N.
For 1 drop = 50 ppm, use R-0735G Hydrochloric Acid .6N. 1.

Rinse and fill 25 mL sample tube (#9198G) to 25 mL mark with water to be tested.

NOTE: For results in grains per gallon (gpg), fill to 14.6 mL mark.

2. Add 3 drops R-0638G Phenolphthalein Indicator. Swirl to mix. Sample will turn pink if P alkalinity is present--proceed to Step 3. If no pink color, go to Step 4.
3. If pink, add R-0724 Hydrochloric Acid .12N or R-0735G Hydrochloric Acid .6N dropwise, swirling and counting after each drop, until color changes from pink to colorless. Record drops as P reading. Always hold bottle in vertical position.
4. Add 3 drops R-0637 Methyl Orange Indicator. Swirl to mix. Sample should turn yellow.
5. Continue adding R-0724 Hydrochloric Acid .12N or R-0735G Hydrochloric Acid .6N dropwise, swirling and counting after each drop, until color changes from yellow to orange (salmon pink). Record total drops (Steps 3 & 5) as M reading. Always hold bottle in vertical position.
6. If R-0724 Hydrochloric Acid .12N is used, multiply P reading by 10. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply M reading by 10. Record as ppm M alkalinity as calcium carbonate. For 14.6 mL sample, record P reading as grains per gallon (gpg) P alkalinity as calcium carbonate. Record M reading as gpg M alkalinity as calcium carbonate. If R-0735G Hydrochloric Acid .6N is used, multiply P reading by 50. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply M reading by 50. Record as ppm M alkalinity as calcium carbonate. For 14.6 mL sample, multiply P reading by 5. Record as grains per gallon (gpg) P alkalinity as calcium carbonate. Multiply M reading by 5. Record as gpg M alkalinity as calcium carbonate. P/T Alkalinity

NOTE:

For 1 drop = 10 ppm, use R-0724 Hydrochloric Acid .12N.
For 1 drop = 50 ppm, use R-0735G Hydrochloric Acid .6N.

1. Rinse and fill 25 mL sample tube (#9198G) to 25 mL mark with water to be tested. NOTE: For results in grains per gallon (gpg), fill to 14.6 mL mark.
2. Add 3 drops R-0638G Phenolphthalein Indicator. Swirl to mix. Sample will turn pink if P alkalinity is present--proceed to Step 3. If no pink color, go to Step 4.
3. If pink, add R-0724 Hydrochloric Acid .12N or R-0735G Hydrochloric Acid .6N dropwise, swirling and counting after each drop, until color changes from pink to colorless. Record drops as P reading. Always hold bottle in vertical position.
4. Add 5 drops R-0645 Total Alkalinity Indicator. Swirl to mix. Sample should turn green.
5. Continue adding R-0724 Hydrochloric Acid .12N or R-0735G Hydrochloric Acid .6N dropwise, swirling and counting after each drop, until color changes from green to pink. Record total drops (Steps 3 & 5) as T reading. Always hold bottle in vertical position.
6. If R-0724 Hydrochloric Acid .12N is used, multiply P reading by 10. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply T reading by 10. Record as ppm T alkalinity as calcium carbonate. For 14.6 mL sample, record P reading as grains per gallon (gpg) P alkalinity as calcium carbonate. Record T reading as gpg T alkalinity as calcium carbonate. If R-0735G Hydrochloric Acid .6N is used, multiply P reading by 50. Record as parts per million (ppm) P alkalinity as calcium carbonate. Multiply T reading by 50. Record as ppm T alkalinity as calcium carbonate. For 14.6 mL sample, multiply P reading by 5. Record as grains per gallon (gpg) P alkalinity as calcium carbonate. Multiply T reading by 5. Record as gpg T alkalinity as calcium carbonate.

Share your knowledge of this product. Be the first to write a review »