#1 in Distributing Specialty
Chemicals & Equipment

Email Us Today Free shipping on
all chemicals (48 states)

  Home > Test Kits, Reagents, Etc. > Taylor Test Kits >

Taylor K-1767
Taylor K-1767
Chlroide Test Kit


 
Taylor K-1767
Our Price: $127.99
Sale Price: $116.99
Savings: $11.00

Quantity in Stock:98


Availability: Usually Ships in 3 to 5 Business Days
Product Code: K-1767
Qty:

Description Shipping & Transit Times
 

Taylor K-1767

ChemWorld is a distributor of Taylor Technologies Test Kits and Reagents. Order directly online and save today.

DROP TEST SALT WATER (1 drop = 20, 40, 100, 200, or 800 ppm)

COMPONENTS: 1 x 4078 Pipet, Graduated, 3 mL (0.5 mL div.), plastic 1 x 5091 Instruction 1 x 9198O Sample Tube, Graduated, 25 mL, plastic w/cap and orange dot 1 x R-0630-C Chromate Indicator, 2 oz, DB 1 x R-0706-C Silver Nitrate Reagent, 2 oz, DB.

NOTE: When sulfite content of sample water to be tested exceeds 10 ppm, the sulfite should be oxidized to prevent interference in test. A 25 mL water sample is first adjusted to the appropriate pH, then 1 mL (or 24 drops) of R-0649 3% Hydrogen Peroxide Solution (sold separately) is added and thoroughly mixed. Continue with the rest of the procedure.

For 1 drop = 20 or 40 ppm Sodium Chloride

1. Select sample size. Desired Equivalence Sample Size 1 drop = 20 ppm 20.5 mL sample 1 drop = 40 ppm 10 mL sample
2. Rinse and fill 25 mL sample tube (#9198O) to desired mark (Step 1) with water to be tested.
3. Add 1 drop R-0630 Chromate Indicator. Swirl to mix. Sample should turn yellow.
4. Add R-0706 Silver Nitrate Reagent dropwise, swirling and counting after each drop, until color changes from yellow to red. Always hold bottle in vertical position.

NOTE: Do NOT add enough R-0706 Silver Nitrate Reagent to give a brown color. First change from yellow to red is the endpoint.
5. Multiply drops of R-0706 Silver Nitrate Reagent by desired equivalence factor (Step 1). Record as parts per million (ppm) salt as sodium chloride.

For 1 drop = 100, 200, or 800 ppm Sodium Chloride

1. Select sample size. Desired Equivalence Sample Size

1 drop = 100 ppm 4 mL sample (2 x 2 mL)
1 drop = 200 ppm 2 mL sample 1 drop = 800 ppm 0.5 mL

2. Using a 3 mL pipet (#4078), add desired sample size (Step 1) to 25 mL sample tube (#9198O). Dilute to 10 mL mark with distilled, deionized, or tap water.
3. Add 1 drop R-0630 Chromate Indicator. Swirl to mix. Sample should turn yellow.
4. Add R-0706 Silver Nitrate Reagent dropwise, swirling and counting after each drop, until color changes from yellow to red. Always hold bottle in vertical position.

NOTE: Do NOT add enough R-0706 Silver Nitrate Reagent to give a brown color. First change from yellow to red is the endpoint.
5. Multiply drops of R-0706 Silver Nitrate Reagent by desired equivalence factor (Step 1). Record as parts per million (ppm) salt as sodium chloride.

Share your knowledge of this product. Be the first to write a review »